Laboratory nitrogen generator Cruise Series(Cruise I)

Cruise Series: In order to meet the demand of multi-channel nitrogen and air, we launched the cruiser series. The cruiser series is divided into three sub-series, respectively, cruise I, mainly aimed at AB liquid products more than 8050 series of liquid products, there is 1 channel nitrogen 1 channel air; cruiser III, mainly a number of liquid products, and their own compressed air station customers, good scalability, nitrogen flow range, can provide from 60 L/min -240L/min..

Cruise I Features: Good reliability: multi-channel design, non-interference, independent and stable; Operational good :1 channel nitrogen ,2 channels air flow control alone, pressure control alone, easy to operate; Good traceability: running time cumulative record, can trace the overall running time.

Cruise I technical parameters

 

CruiseA212

 

CruiseA224

 

CruiseA218

Curtain Gas (nitrogen) 12L/min 60-80psi 24L/min 60-80psi 18L/min 60-80psi
Source Gas (air) 24L/min 110psi 48L/min 110psi 26L/min 110psi
 Exhaust Gas (air) 8L/min 60psi 16L/min 60psi 25L/min 60psi
Nitrogen purity 97%~99.5% 97%~99.5% 97%~99.5%
Dew dew point -50℃ -50℃ -50℃
Noise  52 db  52 db  52 db
Electrical requirements 230V50/60Hz 230V50/60Hz 230V50/60Hz
Power consumption 0.75 kw 1.45 kw 1.45 kw
phthalate None None None
Cleanliness <0.01 u <0.01 u <0.01 u
Residual liquid None None None
Maximum altitude 2000 m 2000 m 2000 m
Working environment 5-35℃/41-95℉ 5-35℃/41-95℉ 5-35℃/41-95℉
Boot time 30 min 30 min 30 min
Nitrogen interface dimensions 6 6 6
Shape dimensions 60*80*100 60*120*140 60*120*140
Weight kg 135 185 kg 185 kg
Features: Good authenticity: the first domestic use of nitrogen purity on-line detector, real-time monitoring of nitrogen purity. Good reliability: adopt multi-channel design, not dry each other, independent and steady; Operational good :1 channel nitrogen ,2 channels air flow control alone, pressure control alone, easy to operate; Good traceability: running time cumulative record, can be traced to the overall running time. Applicable liquid type AB:     QTOF series 6600,5600+,4600, X-500R; QTRAP Series SQ6500SQ5500SQ450040003200; Triple Qua Series STQ6500550045003500.